Calibration standard · side cutaway
Pull: cable leads, probe travels left to right
Eddy current into the pageOut of the pageCrowding around a flawField reach (no current in air)Brightness = current density, fading with depth
Impedance · abs
Coil A alone
Amplitude · strip chart
Diff verticalDiff horizontalAbs vertical
Measurements
Each frequency normalized: 4×20% FBH = 4.0 V (diff), 100% TW = 40°
| Indication | Diff Vpp | Diff angle | Depth from curve | Abs Vpp |
Things to try
- Stop just past the 100% hole. The first loop is drawn. Keep pulling and the second loop forms on the other side of null.
- Stretch the wear scar past the coil spacing. The differential splits into two small signals at the scar's ends while the absolute stays one hump.
- Park coil A under the 60% hole. The eddy currents can't cross the hole, so they crowd through the thin ligament under it and glow white.
- Drop to 25 kHz and pull past the ID groove and the support ring. The ID groove nearly disappears and the support takes over the screen.
Frequency effects
400 kHz · Inconel 600
- Std. depth of penetration
- —
- Wall (0.049") ÷ depth
- —
- Field left at OD
- —
- Phase spread 100% → 20%
- —
- ID groove vs 4×20%
- —
- Support vs 4×20%
- —
- Lower frequency goes deeper. Standard depth of penetration grows as 1/√frequency, so the field reaches the OD and past it.
- Deeper field, less phase spread. Phase lag builds with depth measured in penetration depths. When the whole wall is a small fraction of one penetration depth, flaws at different depths land at similar angles and the curve bunches up.
- High frequency favors the ID. Current crowds the inside surface, so ID flaws read strong and the angles spread out for sizing.
- Low frequency favors what's outside. Supports, deposits and OD conditions grow, which is why the low channel is for locating supports and feeding the mix, not sizing.